Small angle x-ray scattering

Bruker Nanostar

Description: Minimum booking hours: 1 hours, Sample characterizing temperature Range: 298 K, Category: Resolve the size and shape of macromolecules, determine pore sizes, characteristic distances of partially ordered materials.

(I) Low Angle: Q-Range: [0.009 to 0.21 Å inverse]. Size of the particle in diameter: 70 nm to 3 nm.
(II) Medium Angle: Q-Range: [0.01 to 0.34 Å inverse]. Size of the particle in diameter: 63 nm to 2 nm.
(III) High Angle: Q-Range: [0.05 to 0.72 Å inverse]. Size of the particle in diameter: 12 nm to 0.8 nm.

Specifications

Rotating anode copper source
Maximum operating power: 45 kV 100 mA
Detector: HI-STAR
Entire path of X-ray beam from source to sample and detector and kept under vacuum to reduce and air scattering.